This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Neptune 2 Waste Management System - Rover (2014-01-24)

Starting date:
January 24, 2014
Posting date:
March 18, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38299

Recalled Products

Neptune 2 Waste Management System - Rover 

Reason

The purpose of this correction is to upgrade the Neptune 2 Waste Management System to the new approved specifications. Previously, Stryker had received reports that the Neptune 2 had been connected directly to a chest tube, postoperatively which resulted in fatalities and injuries. The new approved specifications include a change in technology, a control mechanism change as well as a modification to the canister vacuum level regulator (from mechanical to electrical). The Neptune 2 has been re-designed to address the hazards and harms outlined in previously recalls.

Affected products

Neptune 2 Waste Management System - Rover 

Lot or serial number

All lots

Model or catalog number

0702-001-000

Companies
Manufacturer
Stryker Instruments
4100 East Milham Ave.
Kalamazoo
49001
UNITED STATES